Comparison of the precision of the working length of four electronic apical locators with stainless steel files and titanium nickel in mesiovestibular ducts of first lower molars.

Authors

DOI:

https://doi.org/10.21142/2523-2754-0803-2020-028

Keywords:

electronic apex locator, stainless steel and nickel titanium files, Proper Pixi, Root ZX mini, Woodpex III, Raypex 6, working length

Abstract

Objective: The purpose of this study was to compare the precision of the working length of Root ZX mini, Proper Pixi, Woodpex III and Raypex 6 using VDW C-Pilot No. 15 with Maillefer nickel titanium No. 15 and stainless steel files in mesiovestibular ducts of first lower molars. Materials
and methods: This was an experimental in vitro study, including 35 mesiobuccal canals of lower first molars. The real length of 4 electronic apical locators, Root ZX mini, Proper Pixi, Woodpex III and Raypex6, was determined with a digital Vernier caliper, and electronic measurements
were recorded using stainless steel and nickel titanium files. Results: There were no statistically significant differences between the real and the electronic lengths of the different tools studied (p <0.05). Conclusions: Statistically significant differences were obtained when comparing the
precision of Root ZX mini, Proper Pixi, Woodpex III and Raypex 6 with stainless steel files, while the precision of nickel titanium files was not affected.

Downloads

Download data is not yet available.

Downloads

Published

2020-12-10